Miniaturized Genotyping Reactions Using the Echo Liquid Handler

The Echo Liquid Handler enables low volume genotyping reactions in a 384-well format with as little as 500 nL total reaction volume. Precise and accurate drop placement eliminates cross-contamination, while offering the flexibility to transfer from any well of a source plate to any well of a destination plate. Cost benefits from reduction of tip waste and decrease in reagent consumption can be realized without sacrificed data quality.

  • Miniaturize PCR to save reagent cost and waste
  • Increase throughput with lower volume 384-well PCR
  • Contamination-free reagent and sample transfer
  • Eliminate pipette tips and washing to save consumables cost and waste

Miniaturized Genotyping Reactions

Components Volume (nL/rxn) Volume (nL/rxn) Volume (nL/rxn) Volume (nL/rxn) Final Conc. Final Rxn Conc.
Water 712.5 536 356.25 78.125 1.5X 1.125X
5X DNA Probes Master 450 337.5 225 112.5 1.5X 1.125X
Primer Mix (20X) 112.5 84.5 56.25 28.125 1.5X 1.125X
Genotyping Probe, 10X 225 167 112.5 56.25 1.5X 1.125X
Total MM Volume Dispensed 1500 1125 750 375 1.5X 1.125X

DNA Plasmid Dispensed
WT Plasmid 500 375 250 125
Heterozygote 500 375 250 125
Mutation 500 375 250 125
Total Reaction Volume 2000 1500 1000 500

  • Miniaturize PCR as low as 500 nL
  • Simple assay setup allows for varying conditions within one microplate

Echo 555 Liquid Handler

Echo 555 Liquid Handler

Figure 1 Quantification of wildtype heterozygote

Figure 1: (Right) Quantification of wild-type, heterozygote and mutant in 1.0 µL reaction volume

Zero Cross-Contamination Results

Zero cross-contamination schematic for alternating

Zero cross-contamination schematic for alternating WT, HET, and MUT plasmid plate.

Positive-Negative Call by the LightCycler 480

Positive/Negative Call by the LightCycler 480. Red circles indicate positives. Green circles indicate negatives. There were no positive calls for mutant genes indicating a 100% call rate and zero cross-contamination.

Zero cross-contamination with multiple transfer steps

Reactions were dispensed in 2000 nL final reaction volumes in a two step process

Reactions were dispensed in 2000 nL final reaction volumes in a two step process utilizing the Echo Liquid Handler. DNA Probes Master mix including the primer mix and genotyping probe was dispensed, followed by alternating plasmid DNA (wild-type (WT), heterozygote (HET), and mutation (MUT)) in an alternating pattern, shown above.

  • Starting with only a few source wells, the Echo Liquid Handler can transfer from any well to any well with zero cross-contamination.
  • Tipless transfer of reagents can occur without adherence to a format imposed by traditional liquid handlers.
  • Zero cross-contamination with multiple transfer steps of primer, probe, mastermix and samples.

The Echo 500 Series Liquid Handlers revolutionize liquid transfer by using acoustic energy to eject fluids. The Echo Liquid Handler is completely touchless—no tips or nozzles, and no material contacts the same as it moves from source to destination. Low-volume, accurate and precise transfer at volumes of 2.5 nL and up allow for assay miniaturization to previously unattainable volumes. Acoustic transfer is compatible with a wide range of fluid types, enabling high-throughput total PCR assay assembly for applications including genotyping, gene expression and DNA sequencing. Transfer of samples from any well of a source to any destination further minimizes sample consumption and simplifies assay setup and execution.