絶対数測定用標準粒子

Total absolute cell count and percent lymphocyte values are usually obtained using the number of white blood cells and lymphocytes counted on an automated hematology instrument multiplied by the percentage of positivelystained cells measured on a flow cytometer. This Standard (Indirect) Method, however, is associated with wide interlaboratory variation.1-4 Flow-Count Fluorospheres is an efficient alternative to the Standard (Indirect) Method for absolute counts and provides this measure directly using the single platform of flow cytometric immunophenotyping analysis. When identical volumes of a biological specimen and Flow-Count Fluorospheres are present, a ratio of cells in the specimen to fluorospheres is established. The cells of interest and the fluorospheres are then counted on the flow cytometry system.

Explore Absolute Counting Beads Models

関連製品

AQUIOS Flow-Count標準粒子

AQUIOS Flow-Count標準粒子は、AQUIOS CLフローサイトメーターで生体サンプルに含まれる絶対白血球数を直接測定するための蛍光マイクロスフィア試薬です。Flow-Count標準粒子は、フローサイトメーターでの絶対数測定に使用する蛍光マイクロビーズ懸濁液です。各標準粒子には、488 nmでの励起時の蛍光粒子放射範囲が525 nmから700 nmである色素が含まれています。これらの粒子のサイズと蛍光強度は一定であり、アッセイ値は絶対数を直接測定できるように設定されています。

技術資料

資料が見つからない場合は、 こちらの検索ページでご確認ください